FC300 Operating Specifications

  • Operating fluid pressure range (LP): 0 to 172 kPa ( 0 to 30 psig) @ 80°C (176°F)
  • Operating fluid pressure range (HP – 10″ only): 0 to 516 kPa ( 0 to 45 psig) @ 80°C (176°F)
  • Operating fluid temperature range: 1 to 80°C (33 to 176°F)
  • Maximum Pressure differential on tubing: 35 kPa (0 to 5 psid) @ 80°C (176°F)
  • Negative pressure differential will collapse tubing and must be avoided
  • Storage temperature range: -30 to 60°C (-22 to 140°F)

FC300 Materials of construction

  • Housing / Shell – GE Noryl
  • End caps – GE Noryl
  • Seals – EPDM (peroxide cured)
  • Tube bundle headers – Thermoset polyurethane
  • Membrane tubing – Nafion™ polymer

IMPORTANT – WARNING! Due to differing rates of expansion, metal fittings are not recommended. ABS or Nylon fittings are recommended. Additionally, DO NOT USE a thread sealant compound when attaching fittings to Perma Pure humidifiers, it is chemically incompatible with the assembly, leading to leaks (and voiding the warranty). Use Teflon tape instead.

FC300

FC 300 Dimensions

Model Number A B
Nom.
C D Flow Rate (slpm)
Gas-to-Gas Water-to-Gas
Approach Dew Point Temp.* Flow vs Dewpoint
4 deg C
6 deg C 9 deg C
FC300-1660-7LP 12.10″ 7″ 5.60″ 2.1″ 121 slpm 291 slpm 387 slpm
FC300-1660-10LP/HP 15.10″ 10″ 8.60″ 5.1″ 173 slpm 415 slpm 553 slpm
FC300-1660-15LP 20.10″ 15″ 13.60″ 10.1″ 259 slpm 623 slpm 830 slpm N/A

* The performance information for the gas-to-gas units are expressed as the flow rate of air (slpm) in relation to the approach dew point temperature. The approach temperature is defined as the difference between the cathode exhaust stream entering the humidifier and the humidifier exhaust stream going to the stack. For example, with a FC300-1660-7LP, if the cathode exhaust dew point entering the humidifier is 67° at 291 slpm, the humidifier generates an output dew point of 61°C (approximately a 6°C difference from the cathode exhaust dew point).